m/z 5 – 1250
≤0.1 amu per 24 h
≥500 MRM/s
IDL <2.0fg for reserpine, ESI+
Nitrogen‑Only Operation



info@bikaicorp.com
TQ 1621
BIKAI Triple Quadrupole LC-MS/MS · Models A and B
BIKAI TQ 1621 triple quadrupole LC‑MS/MS achieves femtogram‑level instrument detection limit, operating purely with nitrogen gas without requiring argon gas cylinder. Equipped with a 180° curved collision cell with axial linear acceleration, it eliminates neutral particles from ion beam and restricts MRM crosstalk down to 0.01%, preserving excellent sensitivity even for complex real‑world sample matrices.
The complete system is delivered with integrated PSA nitrogen generator and oil‑free dry fore pump with 5‑year maintenance‑free service life. It removes hassle of argon supply contract, gas cylinder management and regular oil changes. The instrument is supplied as a fully functional working system out‑of‑the‑box.
| Sensitivity specification | TQ 1621A | TQ 1621B |
| S/N, ESI+ MRM (1 pg reserpine, RMS) | ≥1,500,000:1 | ≥2,000,000:1 |
| IDL, ESI+ (10 fg reserpine, n≥8, 99%) | ≤4.0 fg | ≤2.0 fg |
| IDL, ESI− (10 fg chloramphenicol, n≥8, 99%) | ≤6.0 fg | ≤4.0 fg |
| Everything else | Identical — same ion source, analyser, collision cell, detector, vacuum system and software | |
| Configuration | Spatial tandem triple quadrupole, circular gold-plated rods |
| Mass range | m/z 5 – 1250 |
| Resolution | 0.4 – 2.5 FWHM, adjustable across the full mass range |
| Mass stability / accuracy | ≤0.1 amu per 24 h; ≤±0.1 amu |
| Crosstalk | ≤0.01% |
| Dynamic range | 6 orders of magnitude |
| MRM scan speed | ≥500 MRM/s |
| Reproducibility | Peak area RSD below 1.5%, RT RSD below 0.25% (5 pg tacrolimus, n≥8) |
| Collision cell | Q2, 180° curved linear acceleration; high-purity N₂, no argon required |
| Detector | Discrete-dynode electron multiplier |
| Fore pump | Oil-free dry pump ≥90 m³/h, five years maintenance-free |
| Dimensions and weight | 800 × 852 × 590 mm; 208 kg |