m/z 5 – 2000
>10,000 amu/s
S/N Ratio>100:1
Source Swap in Under 30 Seconds
six orders of magnitude



info@bikaicorp.com
SQ 1610
SQ 1610 Single Quadrupole Mass Spectrometer
BIKAI SQ 1610 single‑quadrupole LC‑MS is engineered for outstanding anti‑contamination performance. Instrument downtime is mostly caused by contamination rather than hardware specifications. SQ 1610 addresses contamination risks throughout the full ion‑path: orthogonal spray design, active exhaust system, heated counter‑flow curtain gas paired with a 0.3 mm cone orifice (without consumable capillary), and gold‑plated molybdenum quadrupole. These four independent protective measures together deliver stable mass‑axis performance with mass stability of 0.2 amu over 24‑hour operation.
Delivered as a complete analytical system bundled with a binary UHPLC front‑end, providing seamless liquid‑phase separation and mass detection workflows.
| Analyser | 200 mm circular pure-molybdenum quadrupole, gold-plated |
| Resolution | Unit mass resolution |
| Mass axis stability | 0.2 amu / 24 h |
| Repeatability | Reserpine peak area RSD below 8% |
| Acquisition modes | Full scan, SIM, simultaneous full scan and SIM, alternating scan |
| Detector | Discrete-dynode electron multiplier |
| Vacuum system | Three-stage differential; 40 m³/h rotary; 30 and 60 L/s oil-free multi-port turbo, air-cooled |
| Dimensions | 330 × 350 × 585 mm |
| Data system | Independent acquisition and data-processing software; English, Chinese and Russian interfaces |
Supplied with a binary UHPLC front end
| 18,000 psi | 0.001–5.000 mL/min | 4 °C – ambient | RT+5 – 90 °C |
| Maximum system pressure — 1.7 to 5 µm columns, HPLC and UHPLC on one system | Parallel short-stroke binary gradient pump, four solvent selection, active plunger wash | Cooled autosampler, 4 × 54 vials or 4 × 96-well, carryover below 0.0025% | Forced-air column oven, ±1 °C, four 300 mm columns with optional switching valves |
Have questions or want to learn more? We’d love to hear from you.